SYT-502 syringe pump

Product Code: SYT-502  syringe pump

500ii

 Three kinds of modes: Rate Mode, Time Mode, Body Weight Mode.
 Double CPU, making the process of injection safer and more reliable.
 Accurate and sensitive pressure sensor detecting occlusion pressure accurately.
 Injection time, dose, solution can be checked during the process of injection.
 Precision step motor. There is no pulsating wave even during low-rate injection.
 KVO rate can be selectable according to concrete requirements from 0.1ml/h to 5ml/h.
 Excellent injection operation. The accuracy will be ensured even at small-volume injection.
 Unique injection supervision system ensures the safe injection and reduces the nurse’s workload.
 Accept all standard syringes of 5ml, 10ml, 20ml, 30ml and 50ml. It will self-diagnose and display the syringe size and set parameter automatically in the form of menu.
 Alarm is given when there is malfunction or operation error, and the machine stops running automatically.
 Save the preset volume automatically when there is unpredictable switch off, ensuring the accuracy of injection volume

Applicable syringes All brands of standard syringes: 5ml,10ml,20ml,30ml,50ml Weight <3.0kg
Accuracy ±2% Dimension 380×190×180mm
Injection increment 0.1ml Power supply AC100~240V, 50/60Hz
Occlusion value 40kPa~160kPa Fuse T2AL250V~
Injection rate 5ml: 0.1 ml/h -100ml/h   30ml: 0.1 ml/h -600ml/h

10ml: 0.1 ml/h -200ml/h  50ml: 0.1 ml/h -1500ml/h

20ml: 0.1 ml/h -400ml/h

Maximum power consumption 25W, can run for 4h after being recharged
Battery Rechargeable lithium polymer battery, 7.4V, 1600mAh
Accumulated injection volume 0.1ml~9999.9ml  

Battery recharge

When the pump is connected to the AC power supply, turn on the power and the battery will automatically recharge (about 8~14 hours to recharge fully.)
Modes of operation Selectable: Rate mode, Time mode, Body Weight mode
Time mode Volume limit: 0.1~999.9ml  Time: 1~2000 minutes
Electrical safety

 

Comply with the stipulation of IEC60601-1: 1996
Body weight mode Drug limit:0.1~999.9mg, Solution limit:0.1~999.9ml, Weight: 01~300.0kg, Dose: 0.1~9999.9 mg/kg/h, ug/kg/min
Electromagnetic compatibility Comply with the stipulation of IEC60601-1-2: 2006
Displayed information Rate, volume limit, accumulated injection volume, syringe size, pressure value, battery capacity, unit of medical dosage, bed No., AC power indicator, battery charge indicator, drug name, bolus rate Shell material ABS plastic
Classification II Class-BF
Alarm function OVER, NEAR, Occlusion, Low battery, Syringe disengaged, No AC power supply, serious low battery Operating conditions Environment temperature 5°C ~40°C, air pressure 80kPa ~106kPa, relative humidity ≤80%
Storage conditions Environment temperature -40°C~55°C, air pressure 50kPa ~106kPa, relative humidity ≤95%
KVO rate 0.1ml/h-5.0ml/h (selectable) Applied standards IEC60601-1: 2005

IEC60601-2-24: 1998

IEC60601-1-2: 2006

ISO780-1997

ISO15223: 2000

ISO7886-1: 1993

ISO / IEC: Guide No. 37《Consumer Goods Usage Instructions》

Bolus rate 5ml: 100ml/h

10ml: 200ml/h

20ml: 400ml/h

30ml: 400ml/h-600ml/h

50ml: 400ml/h-1500ml/h